Reliable Reproducible Repeatable Representative Regulatory

Home / Products /

5E-TCN2200 Nitrogen/Protein Analyzer

5E-TCN2200 Nitrogen/Protein Analyzer

The 5E-TCN2200 is a powerful Dumas Protein/Nitrogen Analyzer designed for precise nitrogen analysis and protein determination, also known as the “Combustion method.” This analyzer ensures accurate results within minutes by breaking down the sample into its elemental compounds. Substances produced, such as H₂O, O₂, and CO₂, are trapped and removed to ensure the correct detection of nitrogen. It serves as an alternative to the Kjeldahl method for determining nitrogen and protein in various types of samples. The Dumas method is the faster official method, validated by different international organizations such as AOAC, AACC, ISO, OIV, ect.

Product Category: Analytical Instruments, Agriculture & Environment, Coal, Metal Ore Mining, Nitrogen/Protein Analyzer.


Features

  • Application Versatility: The analyzer is capable of performing precise nitrogen analysis and protein determination using the Dumas method, also known as the “Combustion method.” This method ensures accurate results in just a few minutes by breaking down the sample into its elemental compounds. It is an alternative to the Kjeldahl method and can be used for various samples, including food, feed, soil, and liquids
  • High Accuracy: The analyzer ensures precise results by trapping and removing substances such as H₂O, O₂, and CO₂ to ensure correct nitrogen detection
  • Compliance with Standards: The analyzer conforms to several international methods and standards, including AOAC, AACC, ASBC, ISO, and OIV
  • Fast Analysis Time: The analysis time is approximately 5 minutes, depending on sample combustion conditions
  • Large Sample Capacity: The stackable auto loader can handle up to 140 samples by 4 layers, allowing for efficient processing of multiple samples
  • Repeatability: The analyzer has a repeatability of Nitrogen (N ad) ≤ 0.05%, ensuring consistent and reliable results
  • Wide Measurement Range: The measurement range for nitrogen is 0.04mg to 50mg, accommodating a variety of sample sizes and types
  • High-Temperature Capability: The resistance furnace (main furnace and furnace rear) can reach a maximum temperature of 1050°C, allowing for effective combustion of samples
  • Gas Requirements: The analyzer requires high-purity gases, including helium, oxygen, and nitrogen or compressed air, to ensure accurate analysis
  • User-Friendly Design: The analyzer features a modern design with improved navigation and enhanced functionality to provide a better user experience.

Application

The CKIC Nitrogen Protein Analyzer is a powerful Dumas Protein/Nitrogen Analyzer designed for precise nitrogen analysis and protein determination. Known as the “Combustion method,” it ensures accurate results in just a few minutes by breaking down the sample into its elemental compounds. Substances such as H₂O, O₂, and CO₂ are trapped and removed to ensure correct nitrogen detection. This method serves as an alternative to the Kjeldahl method for determining nitrogen and protein content in various samples, including food, feed, soil, and liquids. The Dumas method is the faster official method, validated by several international organizations such as AOAC, AACC, ASBC, ISO, and OIV.

Reliable, Automated, and Cost-Effective

The 5E-TCN2200 is a powerful Dumas Protein/Nitrogen Analyzer is designed for efficiency, accuracy, and ease of use, making it an essential tool for laboratories requiring high-precision analysis with minimal downtime.

📩 Contact us for more information or to request a demonstration.

Datasheet

SpecificationDetails
Analysis Time7 minutes, depending on sample combustion conditions
Sample LoaderStackable auto loader, up to 140 samples by 4 layers
RepeatabilityCarbon (C ad) ≤ 0.5%, Hydrogen (H ad) ≤ 0.15%, Nitrogen (N ad) ≤ 0.08%
Sample MassUp to 1000mg, depending on sample matrix
Temp. Resolution1°C
Gas RequiredHelium, 99.99%, 0.25 ± 0.01 Mpa; Oxygen, 99.99%, 0.25 ± 0.01 Mpa; Nitrogen or compressed air, 0.25 ± 0.01 Mpa
Measurement RangeCarbon: 0.02mg-150mg;Hydrogen: 0.1mg-12mg;Nitrogen: 0.04mg-50mg
Furnace TypeResistance furnace (main furnace and furnace rear), max. temp 1050°C
Power SupplySingle phase, AC220 ±10%, 50/60Hz, 5.5kW
Net Weight110kg
Dimensions (L×W×H)690mm×750mm×720mm